Model: CNME011106
Technical Specification:
Main Frequency: 512KHZ
Power Supply: 110V/60HZ; 220V/50HZ
Weight: 6.5kg
Dimension: 28*38*40
Power Readouts:
Agree with actual power into rated to within ±20% or 5W, whichever is greater, Dosage Error
High Frequency Leakage:
Monopolar: <150mA
Bipolar: <60mA
Low Frequency Leakage:
Normal Polarity, Intact Chassis Ground: <10μA
Normal Polarity, Ground open< 50μA
Reverse Polarity, Ground Open< 50μA
Sink Current, 140V Applied, All inputs < 50μA
Technical Function:
REM (Return Electrode Monitoring)
Return electrode contact quality monitoring system (REM). The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected, which minimize the danger of burning incidents.
Compatible with Other Equipment
Optimum Smoke Evacuation System
Multilanguage Available
Language options: English, Spanish, Portuguese, Turkish and etc
Application of Use
General Surgery; Dermatology; Vascular Surgery; Obstetrics and Gynecology ORL/ENT; Minimally Invasive Surgery (MSI)
Cerebral surgery; Neurosurgery, Orthopedics & Plastic Surgery.