General Features:
-New concept and more friendly software
-New techniques and spare parts used for much better quality and function.
-Automatic washing station for both sample and reagent probe.
-Liquid level detection and collision protection for both sample and reagent probe.
-Reliable cooling system for reagent chamber.
-Powerful reaction cuvette washing system.
-ISE module and barcode scanner (optional).
-Lower waste and more environment friendly.
-New concept test control system available (on request).
Technical Specification:
Assay Model: Random access with direct reading system
Method: End point, Kinetic, Fixed time, Immunoturbidimetry, 1-2 reagent, Multistandard, reagent/serum blank, etc
Assay Item: 40 assay
ISE mould (Na+ , K+ , Cl- /Li+) available on request
Throughput: 1 Reagent: 240 tests/h(max)
2 Reagent: 150 tests/h
With ISE: +120 tests/h
Test Menu: TBILI r-GT/GGT CHO Ca APOB-1/B CHE
D.BILI ALP/AKP TG CL APOB CREA
TP UREA CK P CK-MB Mg
ALB CREA LDH C ASO AFU
Lactic Acid ALT/GPT UA α-HBDH CO2 FMN
LP(a) AST/GOT GLU AMY LDL-C HDL-C
CHE LA TBA ADA HS-CRP PA
Fe Cu Zn C3 C4 LGG
LGM LGA and drug, toxicity, etc
Sample position: 60
Including standard, control and stat; either serum cup or primary tube available
Sample Volume: 1~100ul, 0.1ul/step
Reagent Position: 40 or 80(optional)
Reagent Volume: R1:1~400ul, 1ul/step, R2: 1~400ul, 1ul/step
Reagent/Sample probe: With liquid level sensor, collision sensor, Teflon coating and automatic washing station.
Reaction Time: 0~999 seconds
Cuvette Washing High quality uv-transmitted plastic cuvette. Quartz glass cuvettes available on request.
Cuvette Washing: 8 channels automatic washing system for reaction cuvettes
Probe and Mixer Washing: Independent washing station for each sample/reagent probe and mixer
Alarm: Automatic alarm for malfunction (eg: collision, reagent/sample insufficiency, full waste container, etc)
Calibration: Liner, Non-liner, Multi standard, K factor
Sample Dilution/Retest: Sample can be diluted and retested if results are out of range or sample is insufficient. Also with calibrator auto diluting function
Wavelength: 340~810nm
Light Source: Halogen lamp
Absorbance Resolution: 0.0001A
Absorbance Accuracy: ±0.0003A(0~2.5A)
Repeatability (CV%): ≤2%
Quality Control: Levy Jennings quality control program at 3 levels
Bar Code: Optional
Printer: Multiple report formats available
Test Control Function: New card control system available on request
Language: English, French, Spanish, German, Russian, etc
System: Windows XP or Vista; Windows 7
Power Supply: 220V/50Hz or 110V/60Hz (optional)
Dimension: 85*72*53(cm)
Weight: 85kg