Home  >  Products  >  Emergency and Clinics Apparatuses  >  Defibrillator
Automated External Defibrillator Machine
  • Automated External Defibrillator Machine
Automated External Defibrillator Machine

Get Quotes

Download Full Catalogue

Automated External Defibrillator Machine

Model: CNME011304

Technical Specification:

Physical

 

Dimensions:

210 mm (w)x 286 mm (d)x 79.5 mm(h)

Weight:

Main unit (Pro): 2.9 kg(without battery)

Main unit (Public):2.7 kg (without battery)

Rechargeable battery:0.51 kg

Disposable battery:0.42 kg

Environmental

 

Water Resistance: IP5x

Solids Resistance: IP5x

Temperature:

Operating:0 to 50℃

Storage:-30 to 70℃

Humidity:

Operating/Storage:0 to 95 %,non-condensation

Altitude: Operating/Storage:-381 m to+4575 m

Shock and Vibration:

MIL-SID-810G,Method 516.6

MIL-STD-810G,Method 514.6

EN1789(Medical devices for use in road ambulances)

Bump:

EN1789(Medical devices for use in road ambulances)

Free fall:

EN1789(height of fall:0.75m)

IEC60068-2-32(height of fall:1.5 m)

Safety: EN/IEC 60601-1

Display

 

Type: TFT Color LCD

Dimensions: 7 inch

Resolution: 800 x 480 pixels

Display Waveforms: 1 channel

Wave viewing Time: 6 s

Power

 

Rechargeable Battery( for Pro)

Type: 3000 mAh, 14.8v rechargeable lithium ion battery pack

Charge Time:

Fully charged less than 3 hours, 90% charged less than 2 hours

Capacity Indicator:

5-segment indicator for battery capacity evaluation

Capacity (new fully charged battery):Minimum 300 times 200 J discharge or

200 times 360 J discharge

Disposable Battery

Type: Li/MnO2

Capacity (new fully charged battery):

Minimum 300 times 200 J discharge or 200 times 360 J discharge

(The same to rechargeable battery)

Shelf Life: 5 years from date of manufacture

Standby Life: 4 years

Pads

Active Surface Area:

12.8"(83 cm²)each (adult)

6.7(43 cm²) each(pediatric)

Cable Length: 82.6",+4(210 cm,+10 cm)

Use-by Date:

36 months from date of manufacture(adult)

18 months from date of manufacture(pediatric)

Data storage

 

Events: Up to 1000 events

Waveform Storage:

Up to 8 hours of consecutive ECG waveform

Voice Recording: Max.3 hours

Data Export: Through USB flash memory

Defibrillator

 

Waveform:

Biphasic truncated exponential waveform,

with impedance compensation

Energy Accuracy:

±2 Jor 15%, whichever is greater,

Shock Delivery:

Via multifunctional defibrillation electrode pads

Patient lmpedance: 25 to 200 Ohm

Manual Mode ( for Pro)

 

Output Energy:

1,2,3,4,5,6,7,8,9,10,15,20,30,50,70,100,150,170,

200,300,360J

Charge Time:

Less than 5 seconds to 200 J with a new fully charged battery

Less than 8 seconds to 360 J with a new fully charged battery

Synchronous Cardioversion:

Energy transfer begins within 60ms after the QRS peak

AED Mode

Output energy: User configurable

AED shock series:

Energy level for adult: 100 to 360 J, configurable

Energy level for pediatric: 10 to 100 J, configurable

Shock series: 1,2,3, configurable

Default configuration meets 2015AHA/ERC Guidelines

Quick shock:

Typically analyzes and charges in <8 seconds from the end of "Stop CPR" prompt

ECG Monitoring ( for Pro )

 

Lead Type: 3 leads ECG,PADS

Lead Selection: l,ll,lll Pads

Heart Rate Display:

Adult:15 to 300 bpm

Pediatric:15 to 350 bpm

Resolution:1 bpm

ECG Size:

AUTO,1.25 mm/mV(x0.125),2.5 mm/mV(x0.25),

5 mm/mV(x0.5),10 mm/mV(x1),20 mm/mV(x2),

40 mm/mV(x4)

Sweep Speed: 25 mm/s

Automated and User-activated Self-tests

 

Daily Auto Tests:

Test internal circuitry, energy charging/delivery system, pads, and battery capacity

Battery-insert/User Tests: Upon battery insertion, extensive auto self-tests and

user-interactive test check device readiness

Status Indicators:

Blinking green light indicates ready for use.Blinking red light indicates or/ and audible beep indicates need for maintenance


Ask for Quote Now

Download Our Fullcatalogue

Download Our Full Catalogue