Medical Instruments
Medical Instruments
Home  >  Products  >  Emergency and Clinics Apparatuses  >  Anesthesia Machine
7
  • 7
7

7" TFT LCD Display Operation Room ICU Anesthesia Machine

Model: CNME010326

Feature:
7"TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillogram.
High precision flowmeter, instantly know the fresh gas flow to your patient.
Integrated breathing circuit design, ensure easy operating and keep tidy.
Multiple working modes such as volume control and pressure limit, adapt to wide range patient.
Vaporizer with temperature, pressure, flow compensation and self-lock function, keep safety anytime.
Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions in all aspects,
Pressure time, flow-time oscillogram show in real time.
Vital sign monitor and Anesthetic gas monitor are optional.
ACGO and fast oxygen supply- Emergency use and revival after operation.

Safety:
Three level alarm system, visual and sound alarm information.
With multiple type of alarm, reminder and protection functions.
Advanced power management control technology.
Low O2 pressure alarm and N2O cut-off protection.


Specification:

Ventilation modeIPPV, A/C, SIMV, SIGH, MANUAL
Ventilator parameter range Flowmeter: O2(0.1~10L/min)
          N2O(0.1~10L/min)
          AIR(0.1~10L/min)
Rapid oxygen supply: 25L/min~75L/min
Tidal volume(Vt): 0, 20ml~1500ml
Frequence: 1/min~100/min
I:E: 2:1~1:6
PEEP: 0 cmH2O~30 cmH2O
Pressure triggering sensitivity (PTr): -20 cmH2O~0 cmH2O(Based on PEEP)
Flow trigger sensitivity (FTr): 0.5 L/min ~ 30 L/min
SIGH: 0(off) 1/100 ~ 5/100
Apnea ventilation: OFF, 5s~60s
Pressure limit: 20 cmH2O~100 cmH2O
Monitoring parameter Frequence: 0/min~100/min
Tidal volume(Vt): 0mL~2000mL
MV: 0L/min~100L/min
Oxygen concentration: 15%~100%
OscillogramP-T (pressure - time)
F-T (flow- time)
Alarm and protectionThe AC power failure alarm: Power failure or no connection
Low voltage alarm for backup battery: < 11.3±0.3 V
No tidal volume: ≤5 ml within 6s
High oxygen concentration alarm: 19% ~ 100%
Low oxygen concentration alarm: 18% ~ 99%
High airway pressure alarm: 20 cmH2O~100 cmH2O
Low airway pressure alarm: 0 cmH2O~20 cmH2O
High minute volume alarm: Adult (5 L/min -20 L/min)
Paed (1 L/min -15 L/min
Low minute volume alarm: 0-10 L/min
Continuous pressure alarm: (PEEP+1.5kPa) over 16s
Suffocation warning: 5s-60s no spontaneous ventilation
The maximum limited pressure: <12.5 kPa
Fan error: Show on screen
Oxygen deficit: Show on screen
Working conditionGas source: O2, N2O
Pressure: 280kPa~600kPa
Voltage: 100~240V
Power frequency: 50/60Hz
Packing size Wooden case packing size: L740*800*1460mm
G.W: 98KGS
N.W.: 66KGS


Product Name: 7" TFT LCD Display Operation Room ICU Anesthesia Machine
*
Contact Person:
*
Tel:
*
Company Name:
Fax:
*
E-mail:
*
Country:
*
Your Inquiry:
*
CAPTCHA:
">

chat online