Medical Instruments
Medical Instruments
Home  >  Products  >  Emergency and Clinics Apparatuses  >  Anesthesia Machine
Hospital use Anesthesia machine with high precision vaporizer
  • Hospital use Anesthesia machine with high precision vaporizer
Hospital use Anesthesia machine with high precision vaporizer

Hospital use Anesthesia machine with high precision vaporizer

Model: CNME010306

Applications:

The Anesthesia machine makes a good performance in Intensive Care Units (ICU), Operation room,

Anesthesiology Department and other departments. This model is designed for ease of use, incorporating

basic function and the maximum patient safety in daily anesthesia practice. Professional design for

adult, child and infant inhalation anesthesia and respiratory management, with advanced ventilation

mode. Combine proven ventilation technology with the latest refinements in ergonomics and systems

integration with an advanced, easy to use anesthesia table designed together with experienced experts

to streamline your anesthesia workflow.

Features:

7” TFT LCD screen displays the Ventilation parameters, Alarm information and Oscillogram.

High precision flowmeter, instantly know the fresh gas flow to your patient.

Integrated breathing circuit design, ensure easy operating and keep tidy.

Multiple working modes such as volume control and pressure limit, adapt to wide range patient.

Vaporizer with temperature, pressure, flow compensation and self-lock function , keep safety anytime.

Multiple parameters monitoring interface, make every parameter clear, let users know the patient conditions

in all aspects.

Pressure-time, flow-time oscillogram show in real time.

Vital sign monitor and Anesthetic gas monitor are optional.

Technical Specification:

Ventilation mode
IPPV, A/C, SIMV, SIGH, MANUAL
Ventilator parameter rangeFlowmeter: O2(0.1~10L/min)

                    N2O(0.1~10L/min)

Rapid oxygen supply: 25 L/min~75 L/min

Tidal volume (Vt): 0, 20 mL~1500 mL

Frequence: 1 /min~100 /min

I:E: 2:1~1:6

PEEP: 0 cmH2O~30 cmH2O

Pressure triggering sensitivity (Ptr): -20 cmH2O~0 cmH2O (Based on PEEP)

Flow trigger sensitivity (Ftr): 0.5 L/min~30 L/min

SIGH: 0 (off) 1/100~5/100

Apnea ventilation: OFF,5s~60s

Pressure limit: 20 cmH2O~100 cmH2O

Monitoring parameter

Frequence: 0 /min~100 /min

Tidal volume (Vt): 0 mL~2000 mL

MV: 0 L/min~100 L/min

Oxygen concentration: 15 %~100 %

Alarm and protection

The AC power failure alarm: Power failure or no connection

Low voltage alarm for back up battery: <11.3±0.3V

No tidal volume: ≤5mLwithin 6s

High oxygen concentration alarm: 19%~100%

Low oxygen concentration alarm: 18%~99%

High airway pressure alarm: 20 cmH2O~100 cmH2O

Low airway pressure alarm: 0 cmH2O~20 cmH2O

High minute volume alarm: Adult (5 L/min~20 L/min)

Paed (1 L/min~15 L/min)

Low minute volume alarm: 0~10 L/min

Continuous pressure alarm: (PEEP+1.5 kPa) over 16s

Suffocation warning: 5s~60s no spontaneous ventilation

The maximum limited pressure: <12.5kPa

Fan error: Show on screen

Oxygen deficit: Show on screen

Working condition

Gas source: O2, N2O

Pressure: 280kPa~600kPa

Voltage: 100~240V

Power frequency: 50/60Hz

OscillogramP-T (pressure- time)