Model: CNME040101
Specification:
Dimensions(L*W*H): 1450*550*460
Net Weight: Aprox.60kg
Power Supply: AC 220V, -20%~+10%, 50~60Hz
Water Supply
Pressure range: 0~8.0bar
Temperature range: 0~35℃
Water flow: ≥0.3L/min
Technical Data:
VP | -700~+800mmhg, Accuracy±5mmhg |
AP | -700~+800mmhg, Accuracy±5mmhg |
TMP | -700~+800mmhg, Accuracy±5mmhg |
Blood flow | 0~700ml/Min |
Dialysate Flow | 0~300~500~800~1000ml/Min(adjustable) |
Dialysate Temperature | 30~41℃, Accuracy±0.5℃ |
Dialysate Conductivity | 0~25ms/cm, Accuracy±0,1ms/cm |
UF control | UF rate: 0~6000ml/h Accuracy: ±30ml/h or ± 1% |
Substitution Fluid Flow | 0~39L/h (HDF-online) 0~6L/h (HDF-bi bag) |
Heparin injection | 0~10ml/H, Accuracy±0,1ml/H Bolus: 0~10ml/time Syringe size: 20ml, 30ml, 50ml |
Air Detector | Detecting≥0,02ML/single bubble Detect the bubbles |
Blood leak monitor | Detecting blood leak≥0.35mL/min (HCT 32%) |
Clean Disinfection | Decalcification: Rinse, citric acid or glacial acetic acid Thermal Disinfection: Hot water or citric acid (>93℃) Chemical Disinfection: Sodium hypochlorite or peracetic acid |
Technical Features:
•12.1 Inch LCD + touch screen, Chinses, English etc, multi-language system enables the operation much more convenience.
•Dialysate mixed controlling system enables much more accurate ionic concentration of dialysate.
•The advanced volume balance control system enables accurate control of UF.
•Preset a variety of standard dialysate formula, also the formula can be customize for different needs. The formulas are suitable for different brands of concentrated solution or dialysis powder which are available on market.
•Isolated HP mode, decrease the normal equipment’s self -testing time and dialysate waste.
•Ultra-pure dialysate supply, improve patients’ life quality.
•KT/V calculation evaluates the sufficiency of dialysis.
•Bicarbonate cartridge(optional), avoid the waste of B fluid.
•Back-up power supply for 40 mins after electricity off.
•Blood pressure monitor (optional), can be used for monitoring of the patient’s heart rate and blood pressure.
•Venous level detector, the operation is simple and the contamination can be avoided.
Individuation Treatment:
•Patients’ treatment files can be restored.
•Variety profiles of sodium, UF rate, carbonate, heparin, flow, temperature for individual treatment.
•Convenient process for sequential dialysis(Dialysis←→IUF) and high-low sodium sequential dialysis.
•Single\double needle dialysis.
•Bicarbonate and acetate dialysis.
Safety design:
•Reliability and safety treatments are ensured by double safety monitoring system.
•Forcibly and self-test of hydraulic, blood line and monitoring system ensure the safety and reliability of machine.
•Hydraulic system leakage monitoring system ensure the safety of treatment.
•Ultrasonic and optical monitor on venous line ensures the safety of blood return.
•Daily disinfection modes can be preset, all disinfection condition can be record.
•Monitoring system of AP, VP and TMP enable abnormal alarms more quickly.
We have exported to Indonesia, Philippines, Thailand, Morocco, Tunisia, Pakistan, Spain, Nepal, Bolivia, South Africa, and some European countries.
Please contact with us for full catalog and pages.