Description:
Infant Incubator Mode:
For the provision of a similar maternal uterus environment for the neonate to provide stable and controlled comfortable thermal environment and servo-controlled humidity and oxygen level. It is ideal for use in Infant temperature recovery, resuscitation and other occasions.
Infant Radiant Warmer Mode:
Uniform heat distribution, maintaining an optimal thermal environment for caregiver to provide immediate access to the newborn, it is ideal for use in labor and delivery rooms, NICU, and Operating Room.

Brief introduction:
Incubator Mode: Temperature(Air Mode, Baby Mode), Humidity and Oxygen Concentration Control System;
Infant Radiant Warmer Mode: Radiant Temperature Control System (Pre-warm Mode, Manual Mode, and Baby Mode), Timer;
8 inch large size color LCD screen;
Baby bassinet 360° rotatable from both sides of access doors;
Guide rail column in dovetail shape, easy to mount on accessories;
The air circulation system with outlets in both front and back of baby compartment;
RS-232 connector for data output;
X-ray cassette under the bassinet;
Brushless low noise DC motor;
Stepless adjustable inclination of bassinet;
Stored data displayed in curve with Cartesian coordinate axis;
The height of whole unit is adjustable;
Self-check function with multiple failure alarm indications;
Optional Configuration: Weighing system, Observation lamp


Specification:


Product name 

8 Inch Color LCD Infant Radiant Warmer Neonatal Incubator 

Incubator Mode

Mode of control: Air mode and Baby mode controlled by micro-computer

Air temperature control range: 25°C- 37°C

(override mode 37°C-39°C )

Baby temperature control range: 34°C -37°C

(override mode 37°C -38°C)

Accuracy of skin temperature sensor: ±0.2°C

The temperature fluctuation between incubator temperature and average incubator temperature (under stable temperature condition): ≤0.5°C

Temperature uniformity: ≤0.8°C (mattress is on the horizontal position)

≤1.0°C(mattress is on the maximum tilt position)

Inclination of the bassinet:±12° (stepless adjustable)

Noise inside hood (stable working condition): ≤50dB(A)

Relative Humidity Level

Humidity display range: 0%RH~99%RH

Humidity control range: 0%RH~90%RH

Accuracy of humidity display:±5%RH

Accuracy of humidity control:±5%RH

Oxygen Level

Indication range of oxygen concentration: 0%~99%

Control range of oxygen concentration: 20%-60%

Indication accuracy of oxygen concentration:

±2% (Oxygen concentration setting value is 25% and below)

±3% (Oxygen concentration setting value is 25% and above)

Control accuracy of oxygen concentration:±4%

Infant Radiant Warmer Mode

Control Mode: Pre-warm mode, manual mode and baby mode are controlled by micro-computer

Baby mode temperature control range: 32°C- -37.5°C

Control accuracy (The fluctuation range between Skin temperature and Set temperature): ≤0.5°C

Temperature uniformity of mattress: ≤2°C

Weighing System(Optional)

Weighing Range: 100-8000g

Resolution for weight indication: 1g

Accuracy of weighing indication:±1%

Standard Configuration

Main body (including baby compartment, Base, Controller, I.V.pole and tray), Sensor box, Skin temperature sensor and Vertical height adjustment stand VHA stand), Warmer module.

Optional Configuration

Weighing system, Monitor shelf and Disposable skin temperature sensor, Observation lamp, SpO2, etc.

Failure Alarm

Over-temperature alarm, Temperature deviation alarm, Sensor alarm, Motor fan failed alarm, Power failure alarm, Water tank position alarm, Set alarm, Check alarm, and System alarm, etc.
TransportationStorage Temperature range: -20°C~+55°C (Set temperature should be 3C higher than the environment temperature)
Relative Humidity range: ≤93%
Atmospheric pressure: 500hPa - 1060hPa
Environment RequirementOperating range: 20°C- 30°C(Set temperature should be 3°C higher than the environment temperature)
Relative Humidity range: 30%~75%
Ambient air velocity rate: ≤0.3m/s  


Product Packaging:
Package: The whole unit is packed into one carton
Size: 1230mm* 780mm * 1705mm
Gross Weight: 220Kg