Home  >  Products  >  Emergency and Clinics Apparatuses  >  Infant Incubator
Infant Care Equipment Two Control Mode Emergency Neonatal Incubator
  • Infant Care Equipment Two Control Mode Emergency Neonatal Incubator
Infant Care Equipment Two Control Mode Emergency Neonatal Incubator

Get Quotes

Download Full Catalogue

Infant Care Equipment Two Control Mode Emergency Neonatal Incubator

Model: CNME010705plus


Brief Introduction:

Two control modes: Air mode and Baby mode controlled by micro- computer;

Humidity display and control system;

Oxygen concentration control system;

Set temperature, Air temperature, Baby temperature, Humidity and Oxygen are displayed separately;

“>37°C”temperature set function;

Multiple failure alarm indications;

Embedded integrated sensor box;

Integrated water reservoir in the shape of drawer;

Double wall hood, automatic air circulation device;

Inclination of bassinet is adjustable;

RS-232 connector for data output;

Low noise DC motor.

 

Standard Configuration:
Based on CNME010705, with Oxygen concentration control system

Optional Configuration:
Fixed Neonate Bilirubin Phototherapy Equipment model: with blue fluorescent lamps;
Fixed Neonate Bilirubin Phototherapy Equipment model: with blue L ED;
Vertical height adjustment stand (VHA stand) Weighing system, Disposable skin temperature sensor and SpO2, 

Environmental temperature:
Operating range: 20°C~30°C(Set temperature should be 3°C higher than the environment temperature)
Ambient air velocity: <0.3m/s


Specification:

Product name 

Infant Care Equipment Two Control Mode Emergency Neonatal Incubator

Power requirement

AC220V-230V/50Hz or AC110-120V/50-60Hz, 1000VA

Control Mode

Air mode and Baby mode controlled by micro-

computer

Air temperature control range

25°C - 37°C(Override mode 37°C-39°C)

Baby temperature control range

34°C- -37°C(Override mode 37-38°C)

Accuracy of skin temperature sensor

±0.2°C

The temperature fluctuation range between incubator temperature and average incubator temperature (under stable temperature condition)

≤0.5°C

Temperature uniformity

≤0.8°C(mattress is on the horizontal position)

Inclination of the bassinet

±12°

Noise inside hood

≤45dB(A)

Failure alarm

Over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, sensor box position alarm, low water alarm, tank position alarm and system alarm

Humidity display range

0%RH~99%RH

Humidity control range

0%RH~90%RH

Accuracy of humidity display

±5%RH(environment temperature is

25°C, environment humidity is 45%RH±5%RH)

Indication range of oxygen concentration

0%~99%

Control range of oxygen concentration

20%~60%

Accuracy of oxygen display concentration

±2% (oxygen concentration sets ≤25%)

±3% (Oxygen concentration sets >25%)

Accuracy of oxygen control concentration

±4%


Infant Incubator Product Packaging:

Package: Main body and Cabinet are packed into two cartons
Dimension (Main body): 1085mm x 640mm x 915mm
(Cabinet): 1125mm X 665mm X 730mm
Gross Weight (Main body): 68Kg
(Cabinet): 60Kg 