Model: CNME010101
Why does CNME010101 have good price?
For general ward, IBP, AG, BIS and C.O. monitoring are barely used.
Cnme010101 only reserves six parameters, which optimizes its performance, lowers the cost and fully meets basic monitoring needs.Exceptional functions as well as quality.
It is refined and with superior quality. Designed strictly according to CE standards, it is stable and durable with strong antiinterference ability. We inspects every procedure during R&D and manufacturing as per international standards, and carries out clinical trials in hospitals to make sure each product is highly qualified.
1.Parameters: ECG/HR, RESP, SpO2, PR, NIBP, TEMP;
2.ECG Monitoring Technology allows 13 kinds of arrhythmia analysis, ST segment analysis, and improved accuracy of ECG signal measurement;
3.Digital SpO2 Technology enables accurate measurement under weak perfusion and finger trembling situation;
4. Blood Measuring Technology and the study of hemodynamic features of high-risk patients to guarantee precise BP measurement.
Specification:
·Product Size: 300mm*267mm*132mm
·Net Weight: 2.8kg
·Resolution: 800*600
·Power Supply: AC100-250v, 50/6HZ
·Battery Capacity: 2200mA
·Battery Recharging Time: Maximum 6.5 hours for charging
·Battery working time: 2 hours for continuous working
Technical Specification:
RESP:
Method: impedance method
RR measurement range: Adult: 7-120bpm
Pediatric/Neonate: 7-150bpm
Resolution: 1 turns per minute
RESP Apnea Alarm: 10s-40s
Alarm: Audible and visual alarm
Sweep speed: 6.25\12.5\25mms
Gain selection: X1, X2, X4
ECG:
Lead Type: 5 Lead cable: RA; LA; RL; LL or R; L; N; F; C
3 lead cable: RA; LA; LL or R; L; F
Gain selection: X0.25, X0.5, X1, X2
Sweep Speed: 12.5, 25, 50mm/s
Heart Rate Range: Adult: 15-300bpm
Pediatric/Neonate: 15-300bpm
Resolution: 1 bpm
Protection: Withstand 4000VAC/50HZ voltage in isolation
Against electrosurgical interference and defibrillation
Accuracy: ±1% or ±2bpm(whichever is greater)
Mode: Filter MON Mode: 0.5HZ~40HZ
DIA Mode: 0.05HZ~130HZ
OPE Mode: 1HZ~20HZ
ST Mode: 0.2HZ~40HZ
NIBP:
Work mode: Manual/Automatic/Continuous
Auto measure time: Adjustable(1-480min)
Measurement Unit: mmHg/KPA selectable
Measurement Type: Systolic, Diastolic, Mean
Range of systolic pressure: Adult: 40-270mmHg
Pediatric: 40-200mmHg
Neonate: 40-135mmHg
Range of diastolic pressure: Adult: 10-215mmHg
Pediatric: 10-150mmHg
Neonate: 10-100mmHg
Range of mean pressure: Adult: 20-235mmHg
Pediatric: 20-165mmHg
Neonate: 20-110mmHg
Over-pressure protection: both Hardware and software over pressure protection
Accuracy: Less than ±5mmHg
PR from NIBP: 40-240BPM
Resolution: 1mmHg
Temperature(Dual Channel):
Measurement range: 0-50℃(32 to 122F)
Resolution: 0.1℃
Accuracy: ±1℃(without probe)
Channel: 2 channels, T1, T2, TD(Temperature Difference)
SPO2:
Measurement & alarm range: 0~100%
Resolution: 1%
Accuracy: ±2%(70~100%, Ped, non-motion+)
±3%(70~100%, Neo, non-motion); 0-69% unspecified
Alarm range: 0~100%
Pulse rate: Range: 20~250bpm
Resolution: 1bpm
Accuracy: ±3bpm(non-motion) or ±5bpm(motion)
Alarm range: 20-250bpm