Model: CNME010502
Feature:
3” large LCD screen
Can be used for Neonatal, Child, Adult in ICU/CCU/OR in Neurosurgery
Infusion mode: Rate mode, Time mode, Body weight mode
RS232 supported
Safety Requirement:
High accuracy: ±2%, satisfying neonatal use
Fast start, Anti-bolus, Drug name display, History record, Last therapy
parameter memory
IP34 dust and water protection
Interface: Nurse call/DC input
Technical Specification:
Syringe size | 5, 10, 20, 30, 50/60ml |
Flow rate range | 5ml syringe: (0.1~150)ml/h; 10ml syringe: (0.1~300)ml/h; 20ml syringe: (0.1~600)ml/h 30ml syringe: (0.1~900)ml/h; 50ml/60ml syringe: (0.1~1500)ml/h |
Accumulated volume | 0.1~9999ml |
VTBI | 0.1~999.9ml, increment is 0.1ml; 1000-9999ml, increment is 1ml |
Increment | 0.1~999.9ml/h, increment is 0.1ml/h; 1000~1500ml/h, increment is 1ml/h |
Accuracy | ≤ ±2% |
KVO rate | 0.1-5.0ml/h, adjustable, increment is 0.1ml/h |
Bolus rate | 0.1-1500ml/h |
Purge rate | 150-800ml/h |
Occlusion | Low 300±125 mmHg (40kPa); Medium 525±125 mmHg (70kPa); High 900±180 mmHg (120 kPa) |
Handle | Embedded handle |
Mounting type | Standard: Vertical Pole |
Language | 7, Chinese, English, Spanish, French, Russian, Turkish, Portuguese |
Classification | CF, Class I |
IP Grade | IP 34 |
Please contact with us for full catalog and pages.