Model: CNME010519



Specification:
Dimension: 270mm*140mm*80mm(D x W x H)
Weight: 3 Kg
Power Adaptor rate voltage:
Main Power:~100V-240V,50/60Hz
Internal Battery: 10.8V Rechargeable lithium battery
Power Consumption: 100VA
Maximum Flow Rate: 2100(mL/h)

Essential Performance:

Channel number

2 Channel

 

 

 

Rate Range

2mL Syringe:0.01-100mL/h;

5mL Syringe:0.01-150mL/h;

10mL Syringe:0.01-400mL/h;

20mL Syringe:0.01-600mL/h;

30mL Syringe:0.01-1000mL/h;

50mL Syringe:0.01-2100mL/h;

All the above steps were performed at 0.01ml /h.

VTBI Range

0.01~9999.99mL,at step 0.01mL

Accuracy

±2%

 

 

 

Purge Rate(bolus Rate)

2mL Syringe1~100mL/h;

5mL Syringe1~150mL/h;

10mL Syringe1~400mL/h;

20mL Syringe1~600mL/h;

30mL Syringe1~1000mL/h;

50mL Syringe1~2100mL/h;

All the above step at 1mL/h.

 

KVO Rate

3mL for Rate10ml/h

1mL for 1ml/h ≤ Rate < 10ml/h

KVO Rate = Settinge Rate < 1ml/h

Time setting range

00h00min~99h59minat step 1min

 

Occlusion pressure

Maximum:130kPa ± 30kPaMinimum: 26kPa ±

20kPa,level 9 continuously adjustable,See 7.4.7

for setting of occlusion pressure level

Mode

Eight Modes: R+VR+TV+TR, Drug library,

WeightRelay modeGradient mode.

Syringe size

2mL,5mL,10mL,20mL,30mL,50mL

Battery operating time at intermediate rate

Battery can run continuously for more than 3 hours after being fully charged.

 

Wifi

a)Carry out data transmission through WiFi with

"WiFi debugging" app;

 b)The user access control adopts the account

password system and refuses to log in illegally.

Maximum infusion pressure

160kPa

 

 

Time to activate the occlusion alarm; max bolus (Essential Performance)

 

Minimum rate: Occlusion alarm pressure is

26kPa±20kPa, time is < 1h; Occlusion alarm

pressure is 130kPa±30kPa, time is <3h30min.

Medium rate: Occlusion alarm pressure is 26kPa ±20kPa, time<15min, bolus volume is less than 1mL; Occlusion  alarm pressure is 130kPa±30kPa, time<45min, bolus volume is less than 2mL.

(Create a occlusion point in one meter from outlet of syringe)