Model: CNME010311
Description:
20-1500ml (VCV); Applicable to adult, pediatric, infant
Various modes for different needs of all patients
Anesthesia Agent, CO2 and SpO2 monitoring modules (optional)
10.1 inch TFT screen & big font view for easy observation
Touch screen (optional)
Electronic flowmeters for O2, N2O, Air. High accuracy and easy reading
Highly integrated absorber, autoclavable in 134°C to avoid cross-infection (optional)
The absorber could be automatically heated, to avoid water condensation (optional)
Electronic PEEP
Advanced proportional solenoid technology to realize accurate ventilation control
Auxiliary oxygen flowmeter (optional)
Active AGSS system deal with exhaust gas and reduce pollution, providing more safety protection (optional)
Sliding writing table
Compact design, cute and space saving
Technical Specification:
Specification | Ventilation modes: VCV, PCV, SIMV(V)+ PS, SIMV(P)+PS, Spont(optional), Manual, Standby |
Tidal Volume: 20-1500mL | |
Ventilation frequency: 4-60bpm | |
I:E: 3:1-1:6(increment: 0.5) | |
Ti: 0.1-10s(increment: 0.1s) | |
Inspiratory pause(Tp): OFF, 5%-50% | |
Flow Trigger: 1-15L/min | |
Pressure control: 5-70 cmH2O | |
Pressure support: 5-60 cmH2O | |
Pressure limited: PEEP+5-PEEP+ 70 cm H20 | |
Electronic PEEP: OFF, 4-30 cm H2O | |
Monitoring Volume: VTI, VTE, MV, MVspont Pressure: Ppeak, PEEP, Pplat, Pmean, Pmin Freq, Fspont, I:E, Compliance, Resistance, FiO2 SpO2 Module (optional): SpO2, PR CO2 Module (optional): EtCO2, FiCO2 Anesthetic Agent Module (optional): Anesthetic agent, N2O, CO2, and MACAge | |
Waveform: P-T, F-T, V-T | |
Loop: P-V, F-V, F-P, Optional: Volume-EtCO2 | |
Trend: Ppeak, FiO2, MV, PEEP, VTE | |
Alarm: . Audio and visual alarms with three levels Volume: MV high/low, VTE high/low Pressure: Paw high/low, Negative pressure, Low O2 supply pressure Apnea, Continuous high airway pressure Freq high/low, FiO2 high/low, EtCO2 high/low (optional), SpO2 low (optional) Power: AC power failure, Low battery power, Exhausted battery failure | |
Vaporizer | 2 stations with interlock (Selectatec compatible) |
Agent type: Halothane, Enflurane, Isoflurane, Sevoflurane | |
Filler type: Pour fill, Quik fil(only Sevoflurane) | |
Pneumatic Specification | Gas supply: O2, N2O, Air: 0.28-0.6 Mpa |
Electronic Flowmeter: O2/N2O/Air: 0-l0L/Min | |
Gas system: O2 deficiency alarm. Hypoxic guard system, O2 flush: 35-75L/Min | |
Working mode: Closed, Semi-closed, Semi-open, Open | |
Driven mode: Pneumatically driven and electronically controlled | |
Operating mode: Manual/Vent | |
Physical Specification | Dimensions (HxWxD): 1510*1000*700(mm) |
Communication Interface: RS232 | |
Screen: 10.1 inch TFT screen | |
Net weight: 85 kg | |
Power: 100-240VAC, 50/60Hz,6.5A(Max) | |
Lithium Battery: Minimum 120 minutes |