12.1 TFT LCD Touch Screen Anesthesia Machine/Anesthesia Workstation

Model: CNME010305

Technical Specification:

BASE UNIT
Dimensions (H x W x D)

Trolley version (with breathing circuit): 1403mm×704mm×750mm or 55.2in×27.7in×29.5in
Weight and loadTrolley (without vaporizer and backup cylinder): 100kg or 220.5 lb
Top shelf load: 25kg or 55 lb
Power and battery backupPower input: 100~240 VAC, 50/60 Hz
Batteries and Operation time with fully charged: DC24V, 4.0AH, Min.120minutes
Environmental requirementOperation temperature: 10~+40ºC or 50~104ºF
Operation humidity : ≤80%(non-condensing)
Storage temperature: -20~+60ºC or -4~131ºF
Storage humidity: ≤95%(non-condensing)
ANESTHESIA GAS SUPPLY MODULEGas supply: O2, N2O, AIR; 280kPa - 600kPa
Cylinder yokes: Optional: O2, N2O, AIR
Fresh gas flow indicator: Electronically display and virtual fresh gas flow tubes
Range of fresh gas flow indicators: O2, N2O (0.05~10.0 l/min), AIR (0.05~12.0 l/min)
O2 flush: 25~75 l/min
VaporizerAuxiliary common gas outlet (ACGO): Optional Anesthetic Gas Scavenging System (AGSS): Optional
Agent: Halothane, Enflurane, Isoflurane, Sevoflurane
Installation mode: Selectatec with interlock, optional standby vaporizer parking holder
Filling type: Pour Fill, Key Fill, Quik-Fil
Breathing systemVolume of CO2 absorber : 3L for double-canister or 1.8L for optional single canister
VENTILATOR OPERATING SPECIFICATIONSAPL Range: Spontaneous breathing (SP) -70cm H2O
Material: Autoclavable (except O2 cell and mechanical pressure gauge)
Heating system: 32-40ºC
CO2 bypass: Optional
Ventilator : Electronically controlled, pneumatically driven
Ventilation modes: Manual/spontaneous  Volume control (IPPV)  
Pressure control (PCV)  
Pressure Controlled Ventilation Volume Guaranteed(PCV-VG)  
Synchronized Intermittent Mandatory Ventilation in volume(SIMV-VC)  Synchronized Intermittent Mandatory Ventilation in pressure(SIMV-PC)
Pressure support (PS) with Apnea backup
Control input rangesBreathing frequency (rate): 2~100bpm
Positive end expiratory pressure (PEEP): OFF, 3~30cm H2O
Inspiration/expiration ratio (Ti:Te): 4:1~1:8
Tidal volume (Vt): 20~1,500ml in volume control
Inspiration pause: OFF, 5%~60%
Inspiratory time: 0.2~5.0 s
Inspiratory pressure (PTARGET): 5~70cm H2O
Pressure support level (P): 3~50cm H2O
Pressure limit (Pmax): 10~70 cm H2O
Trigger level: 1~15 l/min
Inspiratory Slope Time (TSLOPE): 0~2s
Compensation: Compliance and Leak compensation, fresh gas compensation, altitude compensation
Ventilator monitoring & alarmMonitoring: Continuous monitoring of inspiratory O2 concentration, breathing frequency, tidal volume, minute volume, peak airway pressure, PEEP, mean or plateau pressure,  CO2 concentration (optional), anesthetic gas concentration (optional)
paramagnetic oxygen sensors (optional)
Compliance: 0~100 mL/cm H2O
Control screen: 12.1 in; TFT color touch screen
Graph Display: Waveform of P-t, F-t, V-t, CO2-t (option), Paw-V Loop, V-Flow Loop
Alarm: MV high/low limit, FiO2 high/low limit, Paw high/low limit, Power failure
Breath Rate high limit
Subatmospheric pressure
Sustained airway pressure  
Apnea alarm, Alarm Silence(120s)


Description:
The comprehensive ventilation modes including PCV-VG, SIMV-PC provide more flexibility to ventilation strategy, suitable for various patients operation including pediatric and adult
Advanced ventilation management, provides a accurate and broad of settings enabling effective care across a wide range of patient sizes and acuity types
Support both low-flow anesthesia and mini-flow anesthesia
Compact breathing system, easy to remove and clean, fully autoclavable, optional CO2 bypass
Electronic flow-meter, precise control and convenient operation
12.1'' TFT LCD with color touch screen, highlighted display
Optional comprehensive gas monitoring include: oxygen (paramagnetic), CO2, AG, etc.
CE marked, meets EU clinical requirement