Home  >  Products  >  Emergency and Clinics Apparatuses  >  Infant Incubator
Newborn Baby Incubator Infant Incubator with Humidity Control System
  • Newborn Baby Incubator Infant Incubator with Humidity Control System
Newborn Baby Incubator Infant Incubator with Humidity Control System

Get Quotes

Download Full Catalogue

Newborn Baby Incubator Infant Incubator with Humidity Control System

Model: CNME010706plus


Brief Introduction:

Two control modes: Air mode and Baby mode controlled by micro-computer;

Humidity concentration display system
Humidity control system;
Set temperature, Air temperature, Baby temperature, Humidity concentration are displayed separately;
">37°C" temperature set function;
Multiple failure alarm indications;
Embedded integrated sensor box;
Integrated water reservoir in the shape of drawer;
Double wall hood, automatic air curtain system;
Inclination of bassinet is adjustable;
RS-232 connector for data output;
Oxygen input interface;
Low noise DC motor.

Standard Configuration:
Based on CNME010706, plus Humidity control system

Optional Configuration:
Fixed Neonate Bilirubin Phototherapy Equipment with blue fluorescent lamps;
Fixed Neonate Bilirubin Phototherapy Equipment with blue LED;
Vertical height adjustment stand (VHA stand), Weighing system, SpO2 and Disposable skin temperature sensor.

Environmental temperature:
Operating range: 20°C~30°C (Set temperature should be 3°C higher than the environment temperature)
Ambient air velocity: <0.3m/s 

 

Specification:

Product name 

Newborn Baby Incubator Infant Incubator with Humidity Control System

Power requirement

AC220V-230V/50Hz or AC110-120V/50-60Hz, 1000VA

Control Mode

Air mode and Baby mode controlled by micro-

computer

Air temperature control range

25°C - 37°C(Override mode 37°C-39°C)

Baby temperature control range

34°C- -37°C(Override mode 37-38°C)

Accuracy of skin temperature sensor

±0.2°C

The temperature fluctuation range between incubator temperature and average incubator temperature (under stable temperature condition)

≤0.5°C

Temperature uniformity

≤0.8°C(mattress is on the horizontal position)

Inclination of the bassinet

±12°

Noise inside hood

≤45dB(A)

Failure alarm

Over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, sensor box position alarm, low water alarm, tank position alarm and system alarm.

Humidity display range

0%RH~99%RH

Humidity control range

0%RH~90%RH

Accuracy of humidity display

±5%RH(environment temperature is

25°C, environment humidity is 45%RH±5%RH)


Infant Incubator Product Packaging:

Package: Main body and Cabinet are packed into two cartons
Dimension: (Main body): 1085mm X 640mm X 910mm
(Cabinet): 1125mm x 665mm X 725mm
Gross Weight: (Main body): 66Kg
(Cabinet): 68Kg